IFM SD9000 Product characteristics
|
Number of inputs and outputs |
Number of digital outputs: 2; Number of analogue outputs: 1 |
Process connection |
threaded connection R 1 1/2 DN40 |
Temperature monitoring |
Measuring range [°C] |
0…60 |
IFM SD9000 Application |
Application |
for industrial applications |
Media |
compressed air |
Note on media |
air quality ISO 8573-1 |
class 141 |
class 344 |
|
Medium temperature [°C] |
0…60 |
Pressure rating [bar] |
16 |
Pressure rating [Mpa] |
1.6 |
IFM SD9000 Electrical data
|
Operating voltage [V] |
18…30 DC; (according to EN 50178 SELV/PELV) |
Current consumption [mA] |
< 110 |
Protection class |
III |
Reverse polarity protection |
yes |
Power-on delay time [s] |
1 |
IFM SD9000 Inputs / outputs |
Number of inputs and outputs |
Number of digital outputs: 2; Number of analogue outputs: 1 |
Outputs |
Total number of outputs |
2 |
Output signal |
switching signal; analogue signal; pulse signal; IO-Link; (configurable) |
Electrical design |
PNP |
Number of digital outputs |
2 |
Output function |
normally open / normally closed; (parameterisable) |
Max. voltage drop switching output DC [V] |
2 |
Permanent current rating of switching output DC [mA] |
250; (per output) |
Number of analogue outputs |
1 |
Analogue current output [mA] |
4…20; (scalable) |
Max. load [Ω] |
500 |
Pulse output |
consumed quantity meter |
Short-circuit protection |
yes |
Type of short-circuit protection |
pulsed |
Overload protection |
yes |
IFM SD9000 Measuring/setting range
|
Measuring range |
20…6835 l/min |
0.3…95.3 m/s |
1.5…410 m³/h |
|
Display range |
0…8200 l/min |
0…114.4 m/s |
0…492 m³/h |
|
Set point SP |
55…6835 l/min |
0.8…95.3 m/s |
3.5…410 m³/h |
|
Reset point rP |
20…6800 l/min |
0.3…94.8 m/s |
1.5…408 m³/h |
|
Analogue start point ASP |
0…5125 l/min |
0…71.5 m/s |
0…307.5 m³/h |
|
Analogue end point AEP |
1710…6835 l/min |
23.8…95.3 m/s |
102.5…410 m³/h |
|
In steps of |
|
Volumetric flow quantity monitoring |
Pulse value |
0.005…4 000 000 m³ |
In steps of |
0.001 m³ |
Pulse length [s] |
0,018…2 |
IFM SD9000 Temperature monitoring |
Measuring range [°C] |
0…60 |
Display range [°C] |
-12…72 |
Accuracy / deviations |
IFM SD9000 Flow monitoring |
Repeatability [% of the measured value] |
± 1,5 |
Accuracy (in the measuring range) |
± (3 % MW + 0,3 % MEW) / ± (6 % MW + 0,6 % MEW); (class 141 / ; class 344; conditions: installation to DIN ISO 2533; installation in pipes: DN40) |
Temperature monitoring |
Accuracy [K] |
± 2; (medium flow in the limit area of the flow measurement range) |
IFM SD9000 Response times |
Flow monitoring |
Response time [s] |
0.1; (dAP = 0) |
Damping for the switching output dAP in steps [s] |
0 – 0,2 – 0,4 – 0,6 – 0,8 – 1 |
Software / programming |
Parameter setting options |
Flow monitoring; quantity meter; Preset counter; Temperature monitoring; hysteresis / window; normally open / normally closed; current/pulse output; display can be rotated and switched off; Display unit; totaliser |
Interfaces |
Communication interface |
IO-Link |
Transmission type |
COM2 (38,4 kBaud) |
IO-Link revision |
1.1 |
SDCI standard |
IEC 61131-9 CDV |
IO-Link device ID |
269 d / 00 01 0D h |
Profiles |
no profile |
SIO mode |
yes |
Required master port type |
A |
Process data analogue |
3 |
Process data binary |
2 |
Min. process cycle time [ms] |
4.1 |
IFM SD9000
Operating conditions
|
Ambient temperature [°C] |
0…60 |
Storage temperature [°C] |
-20…85 |
Max. relative air humidity [%] |
90 |
Protection |
IP 65 |
Tests / approvals |
EMC |
DIN EN 61000-6-2 |
|
DIN EN 61000-6-3 |
|
|
Vibration resistance |
DIN EN 68000-2-6 |
5 g (55…2000 Hz) |
|
MTTF [years] |
227 |
Pressure Equipment Directive |
Sound engineering practice; can be used for group 2 fluids; group 1 fluids on request |
IFM SD9000 Mechanical data |
Weight [g] |
4004 |
Materials |
PBT-GF20; NBR; PC; stainless steel (1.4301 / 304); PTFE; brass coated; aluminium powder-coated |
Materials (wetted parts) |
stainless steel (1.4301 / 304); FKM; ceramics glass passivated; PEEK GF30; polyester; aluminium |
Process connection |
threaded connection R 1 1/2 DN40 |
IFM SD9000 Displays / operating elements
|
Display |
Display unit |
5 x LED, green (Nl/min, Nm³/h, Nm/s, Nm³, °C) |
function display |
1 x LED, green |
switching status |
2 x LED, yellow |
measured values |
alphanumeric display, 4-digit |
programming |
alphanumeric display, 4-digit |
|
Remarks |
Remarks |
MW = measured value |
MEW = Final value of the measuring range |
Measuring, display and setting ranges refer to the standard volume flow according to DIN ISO 2533. |
For information about installation and operation please see the operating instructions. |
|
Pack quantity |
1 pcs. |
Electrical connection |
Connection |
|